FDA Device recall Z-0765-2025
Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527
- FDA Device
- Class I
- ongoing
- Published 2025-01-22
On 2025-01-22 the FDA classified a Class I recall of Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527 by Medline Industries Northfield, citing fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0765-2025 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2024-12-16 |
| Published | 2025-01-22 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Medline Industries Northfield |
| Distribution | US |
Product
Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;
Reason
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Identifiers
| code_info | 1) 60130527 UDI-D:10193489046465(each), 20193489046462(case), Lot Number:0000133032; 2) 65111341 UDI-D:10193489061314(e…1) 60130527 UDI-D:10193489046465(each), 20193489046462(case), Lot Number:0000133032; 2) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000129541; 3) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000134701; 4) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000129544; 5) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000128715; 6) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000133060; 7) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000131657; 8) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135723; 9) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135294; 10) 65071193 UDI-D:10193489060904(each), 20193489060901(case), Lot Number:0000130171; 11) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000135606; 12) 600604422 UDI-D:10193489070224(each), 20193489070221(case) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0765-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.