RecallRadar

FDA Device recall Z-0759-2014

MCA5-2SD - 2MM Diamond Ball, 5.6CM. All lots are included in recall

On 2014-01-22 the FDA classified a Class II recall of MCA5-2SD - 2MM Diamond Ball, 5.6CM. All lots are included in recall by The Anspach Effort, citing the Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0759-2014
ClassificationClass II
Statusterminated
Initiated2012-03-29
Published2014-01-22
Last updated2026-09-13 11:53 UTC
CompanyThe Anspach Effort
DistributionUS

Product

MCA5-2SD - 2MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Identifiers

code_infoAll lots are included in recall

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0759-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.