RecallRadar

FDA Device recall Z-0755-2021

Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discret…

On 2021-01-06 the FDA classified a Class II recall of Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discret… by Roche Diagnostics Operations, citing potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters.….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0755-2021
ClassificationClass II
Statusongoing
Initiated2020-12-08
Published2021-01-06
Last updated2026-09-13 11:55 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001

Reason

Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.

Identifiers

code_infoSerial Numbers: 1925-10; 1931-04; 1931-05;1931-07;1937-10; 1938-01; 1942-10

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0755-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.