RecallRadar

FDA Device recall Z-0755-2013

Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100

On 2013-02-06 the FDA classified a Class II recall of Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100 by Ortho Clinical Diagnostics, citing some VITROS¿ Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0755-2013
ClassificationClass II
Statusterminated
Initiated2012-12-13
Published2013-02-06
Last updated2026-09-13 11:53 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS¿ Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS¿ 3600, VITROS¿ 5600, and VITROS¿ ECi/ECiQ Systems.

Reason

Some VITROS¿ Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS¿ System instead of the intended VITROS¿ Vapor Adsorption Cartridge which may bias results.

Identifiers

code_infoCAT No. 680 0100; the cartridges do not have lot numbers.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0755-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.