FDA Device recall Z-0753-2020
Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5…
- FDA Device
- Class II
- terminated
- Published 2020-01-08
On 2020-01-08 the FDA classified a Class II recall of Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5… by Medicrea International, citing it has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0753-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-06-27 |
| Published | 2020-01-08 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Medicrea International |
| Distribution | AZ, CO, FL, NY, OH, PA, TX |
Product
Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only
Reason
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
Identifiers
| code_info | Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0753-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.