FDA Device recall Z-0752-2017
Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instruments intended for…
- FDA Device
- Class II
- terminated
- Published 2016-12-21
On 2016-12-21 the FDA classified a Class II recall of Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instruments intended for… by Integra York Pa, citing the Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0752-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-10-28 |
| Published | 2016-12-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Integra York Pa |
| Distribution | NC, TX |
Product
Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instruments intended for cutting and grasping tissue in general surgical procedures.
Reason
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
Identifiers
| code_info | Catalog # ST5-182; lot# 355 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0752-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.