RecallRadar

FDA Device recall Z-0748-2015

PKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for cutting, holding, grasping an…

On 2014-12-24 the FDA classified a Class II recall of PKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for cutting, holding, grasping an… by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0748-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Identifiers

code_info
841327 941327 941738 942657 942849 943735 944712 944817 945120 945511 1040540 1041287 1042922 1042923 114…841327 941327 941738 942657 942849 943735 944712 944817 945120 945511 1040540 1041287 1042922 1042923 1144222 1145065 1145807 1145968 1242133 1242453 1246631 1246636 1340557 1341515 1342002 1343420 1343725 1344273 1346242 1346691 1440474 642114

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0748-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.