RecallRadar

FDA Device recall Z-0747-2024

48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R

On 2024-01-24 the FDA classified a Class II recall of 48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R by Smiths Medical Asd, citing there is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0747-2024
ClassificationClass II
Statusongoing
Initiated2023-11-10
Published2024-01-24
Last updated2026-09-13 11:57 UTC
CompanySmiths Medical Asd
DistributionUS

Product

48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R

Reason

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Identifiers

code_infoLot Number: 4383431 4387987 4404698

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0747-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.