FDA Device recall Z-0747-2021
Covidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic…
- FDA Device
- Class II
- completed
- Published 2021-01-06
On 2021-01-06 the FDA classified a Class II recall of Covidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic… by Covidien, citing may contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemost….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0747-2021 |
| Classification | Class II |
| Status | completed |
| Initiated | 2020-11-23 |
| Published | 2021-01-06 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Covidien |
| Distribution | US |
Product
Covidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3S
Reason
May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis
Identifiers
| code_info | Lot Numbers: P0B1142MY, P0C1515MY, P0E0634MY, P0E1353MY, P0C0492MY,P0D1454MYP0E1352MY,P0G0601Y, P0C0493MY UDI: 10884521043145, 20884521043142 |
|---|