RecallRadar

FDA Device recall Z-0746-2020

HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the Hear…

On 2020-01-08 the FDA classified a Class II recall of HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the Hear… by Thoratec, citing excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0746-2020
ClassificationClass II
Statusterminated
Initiated2019-12-02
Published2020-01-08
Last updated2026-09-13 11:55 UTC
CompanyThoratec
DistributionUS

Product

HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the HeartMate 3 Left Ventricular Assist System, REF 106524US, UDI or GTIN: 00813024013297, and 106524, UDI or GTIN: 00813024011712 - Product Usage: Provides power to the System Controller and pump. - Provides power to the System Monitor when it is connected to the Power Module. - Connects the System Monitor to the System Controller for monitoring purposes. - Echoes System Controller alarms. The Mobile Power Unit is for home or clinical use when the patient does not require monitoring using the System Monitor. The Mobile Power Unit is used when the patient is indoors, stationary, or sleeping. The System Controller and the Mobile Power Unit are connected through the Mobile Power Unit patient cable. The cable transfers power from the Mobile Power Unit to the System Controller.

Reason

Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.

Identifiers

code_infoAll serial numbers.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0746-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.