Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010

On 2013-02-06 the FDA classified a Class III recall of Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010 by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0745-2013
ClassificationClass III
Statusterminated
Initiated2012-12-20
Published2013-02-06
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.

Reason

Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.

Identifiers

code_infoSiemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.