On 2016-02-10 the FDA classified a Class II recall of ABX PENTRA Uric Acid CP ref by Horiba Instruments, citing N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoanti….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0743-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-01-11 |
| Published | 2016-02-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Horiba Instruments |
| Distribution | AL, AR, AZ, CA, FL, GA, ID, IL, KS, KY, MD, MI, MN, MS, MT, NC, NE, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV |
Product
ABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma and urine based on the enzymatic determination of uric acid using a chromogenic system in the presence of peroxidase and uricase (Trinder method).
Reason
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Identifiers
| code_info | All Lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0743-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.