FDA Device recall Z-0742-2019
ID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduct…
- FDA Device
- Class I
- terminated
- Published 2019-01-30
On 2019-01-30 the FDA classified a Class I recall of ID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduct… by Draeger Medical, citing incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0742-2019 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2018-12-21 |
| Published | 2019-01-30 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Draeger Medical |
| Distribution | AL, CA, CO, CT, FL, KY, MI, NJ, OH, PA, SC, TX |
Product
ID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of condensate. The standard breathing circuit is equipped with a water trap in each breathing hose limb.
Reason
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
Identifiers
| code_info | All Lots |
|---|
Status history