FDA Device recall Z-0742-2016

ABX PENTRA Triglycerides CP ref

On 2016-02-10 the FDA classified a Class II recall of ABX PENTRA Triglycerides CP ref by Horiba Instruments, citing N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoanti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0742-2016
ClassificationClass II
Statusterminated
Initiated2016-01-11
Published2016-02-10
Last updated2026-09-13 11:54 UTC
CompanyHoriba Instruments
DistributionAL, AR, AZ, CA, FL, GA, ID, IL, KS, KY, MD, MI, MN, MS, MT, NC, NE, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV

Product

ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.

Reason

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Identifiers

code_infoAll Lots