FDA Device recall Z-0741-2023
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MM…
- FDA Device
- Class II
- ongoing
- Published 2022-12-28
On 2022-12-28 the FDA classified a Class II recall of InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MM… by Companion Medical, citing diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0741-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-11-08 |
| Published | 2022-12-28 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Companion Medical |
| Distribution | US |
Product
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.
Reason
Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.
Identifiers
| code_info | InPen Software App for Android, versions 5.6.1.1 and 5.6.2 with InPen System REF/UDI-DI: MMT-105ELBLNA/10862088000313,…InPen Software App for Android, versions 5.6.1.1 and 5.6.2 with InPen System REF/UDI-DI: MMT-105ELBLNA/10862088000313, MMT-105ELGYNA/10862088000320, MMT-105ELPKNA/10862088000337, MMT-105NNBLNA/10862088000344, MMT-105NNGYNA/10862088000351, MMT-105NNPKNA/10862088000368. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0741-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.