RecallRadar

FDA Device recall Z-0741-2022

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Numbe…

On 2022-03-16 the FDA classified a Class II recall of Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Numbe… by Cytocell, citing individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0741-2022
ClassificationClass II
Statusterminated
Initiated2022-01-18
Published2022-03-16
Last updated2026-09-13 11:56 UTC
CompanyCytocell
DistributionUS

Product

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Reason

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Identifiers

code_infoLot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0741-2022%22