FDA Device recall Z-0741-2022
Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Numbe…
- FDA Device
- Class II
- terminated
- Published 2022-03-16
On 2022-03-16 the FDA classified a Class II recall of Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Numbe… by Cytocell, citing individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0741-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-01-18 |
| Published | 2022-03-16 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Cytocell |
| Distribution | US |
Product
Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024
Reason
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Identifiers
| code_info | Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0741-2022%22