FDA Device recall Z-0741-2016

ABX PENTRA Lactic Acid CP ref

On 2016-02-10 the FDA classified a Class II recall of ABX PENTRA Lactic Acid CP ref by Horiba Instruments, citing N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoanti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0741-2016
ClassificationClass II
Statusterminated
Initiated2016-01-11
Published2016-02-10
Last updated2026-09-13 11:54 UTC
CompanyHoriba Instruments
DistributionAL, AR, AZ, CA, FL, GA, ID, IL, KS, KY, MD, MI, MN, MS, MT, NC, NE, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV

Product

ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).

Reason

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Identifiers

code_infoAll Lots