FDA Device recall Z-0741-2016
ABX PENTRA Lactic Acid CP ref
- FDA Device
- Class II
- terminated
- Published 2016-02-10
On 2016-02-10 the FDA classified a Class II recall of ABX PENTRA Lactic Acid CP ref by Horiba Instruments, citing N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoanti….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0741-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-01-11 |
| Published | 2016-02-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Horiba Instruments |
| Distribution | AL, AR, AZ, CA, FL, GA, ID, IL, KS, KY, MD, MI, MN, MS, MT, NC, NE, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV |
Product
ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Reason
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Identifiers
| code_info | All Lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0741-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.