RecallRadar

FDA Device recall Z-0739-2016

ABX PENTRA Enzymatic Creatinine CP ref

On 2016-02-10 the FDA classified a Class II recall of ABX PENTRA Enzymatic Creatinine CP ref by Horiba Instruments, citing N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoanti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0739-2016
ClassificationClass II
Statusterminated
Initiated2016-01-11
Published2016-02-10
Last updated2026-09-13 11:54 UTC
CompanyHoriba Instruments
DistributionAL, AR, AZ, CA, FL, GA, ID, IL, KS, KY, MD, MI, MN, MS, MT, NC, NE, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV

Product

ABX PENTRA Enzymatic Creatinine CP ref. A11A01907 reagent is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine based on an enzymatic method of using a multi-step approach ending with a photometric end-point reaction.

Reason

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Identifiers

code_infoAll Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0739-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.