FDA Device recall Z-0737-2025
epoc NXS Host
- FDA Device
- Class II
- ongoing
- Published 2025-01-01
On 2025-01-01 the FDA classified a Class II recall of epoc NXS Host by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0737-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-11-22 |
| Published | 2025-01-01 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Reason
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Identifiers
| code_info | Siemens Material Number: 11413506; UDI-DI: 00630414605821; Software Version Number: 4.14.9. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0737-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.