FDA Device recall Z-0735-2015
PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and ma…
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and ma… by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0735-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-17 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Endoscopy |
| Distribution | US |
Product
PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Identifiers
| code_info | 941070 941263 941599 942144 943866 944196 1041986 1044516 1045038 1045226 1141655 1143142 1243797 1246161 1343131 1343403 1441956 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Stryker Endoscopy
- PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator
FDA Device · Class II · 2016-05-04 · Stryker Endoscopy - SERFAS 90 degree Energy Probe, Part Number 279-350-101
FDA Device · Class II · 2015-07-01 · Stryker Endoscopy - PKG, 5MM INSERT, MINI METZENBAUM SCISSORS, CURVED, 45CM, P/N 0250080574 Laparoscopic Manual Instruments are intended for cutting, holding,…
FDA Device · Class II · 2014-12-24 · Stryker Endoscopy - PKG, BIPOLAR FORCEPS, MICRO TIP, 33CM, P/N 0250080122 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manip…
FDA Device · Class II · 2014-12-24 · Stryker Endoscopy - PKG, 3MM INSERT, FUNDUS GRASPER, 29CM, P/N 0250282050 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manip…