RecallRadar

FDA Device recall Z-0734-2023

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1…

On 2022-12-28 the FDA classified a Class II recall of 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1… by Fresenius Medical Care Holdings, citing (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interferenc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0734-2023
ClassificationClass II
Statusongoing
Initiated2022-07-27
Published2022-12-28
Last updated2026-09-13 11:56 UTC
CompanyFresenius Medical Care Holdings
DistributionUS

Product

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Reason

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Identifiers

code_infoSoftware issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0734-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.