FDA Device recall Z-0733-2019
LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Pro…
- FDA Device
- Class II
- terminated
- Published 2019-01-23
On 2019-01-23 the FDA classified a Class II recall of LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Pro… by Waldemar Link And Kg Mfg Site, citing the reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not scre….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0733-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-12-07 |
| Published | 2019-01-23 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Waldemar Link And Kg Mfg Site |
| Distribution | US |
Product
LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model¿ SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.
Reason
The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
Identifiers
| code_info | Lot numbers: 1435077 1515029 1718103 1435077 150401/0217 1734168 1726056 1718103 1734168 1515029 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0733-2019%22