PKG, 5MM INSERT, ENDO GRASPER, 45CM, P/N 0250080726 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipul…

On 2014-12-24 the FDA classified a Class II recall of PKG, 5MM INSERT, ENDO GRASPER, 45CM, P/N 0250080726 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipul… by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0733-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, 5MM INSERT, ENDO GRASPER, 45CM, P/N 0250080726 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Identifiers

code_info143498 143553 1044398 1044399 1146774 1343552 1344884 1345231 1345234 1345414 1346544 1346661 1440959