FDA Device recall Z-0730-2025
20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatat…
- FDA Device
- Class II
- ongoing
- Published 2025-01-01
On 2025-01-01 the FDA classified a Class II recall of 20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatat… by Abbott Vascular, citing inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0730-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-11-06 |
| Published | 2025-01-01 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Abbott Vascular |
| Distribution | AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI |
Product
20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatation catheters to create and monitor pressure in the balloon.
Reason
Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
Identifiers
| code_info | UDI-DI: 08717648013591, Lots: 60418466, 60418467, 60418468 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0730-2025%22