RecallRadar

FDA Device recall Z-0729-2022

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm

On 2022-03-16 the FDA classified a Class II recall of HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm by Maquet Cardiovascular, citing outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0729-2022
ClassificationClass II
Statusterminated
Initiated2022-01-14
Published2022-03-16
Last updated2026-09-13 11:56 UTC
CompanyMaquet Cardiovascular
DistributionUS

Product

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. Product Code/REF Number: M002020952070

Reason

Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm

Identifiers

code_infoLot Number: 21B10 S/N: 1318567653 UDI: (01)00384401017608(11)210210(17)260131(21)1318567653GS(10)21B10

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0729-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.