FDA Device recall Z-0729-2022
HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm
- FDA Device
- Class II
- terminated
- Published 2022-03-16
On 2022-03-16 the FDA classified a Class II recall of HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm by Maquet Cardiovascular, citing outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0729-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-01-14 |
| Published | 2022-03-16 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Maquet Cardiovascular |
| Distribution | US |
Product
HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. Product Code/REF Number: M002020952070
Reason
Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm
Identifiers
| code_info | Lot Number: 21B10 S/N: 1318567653 UDI: (01)00384401017608(11)210210(17)260131(21)1318567653GS(10)21B10 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0729-2022%22