S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System

On 2020-01-01 the FDA classified a Class II recall of S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System by Cardinal Health, citing cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0728-2020
ClassificationClass II
Statusongoing
Initiated2019-11-14
Published2020-01-01
Last updated2026-09-13 11:55 UTC
CompanyCardinal Health
DistributionUS

Product

S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.

Reason

Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".

Identifiers

code_info
All Lot Codes. Affected SKUs: N1020ABR, N1020BBR, N1020TBR, N1040ABR, N1040BBR, N1040EBR, N1040TBR, N1060ABR, N1060BB…All Lot Codes. Affected SKUs: N1020ABR, N1020BBR, N1020TBR, N1040ABR, N1040BBR, N1040EBR, N1040TBR, N1060ABR, N1060BBR, N1060EBR, N1060TBR, N1080ABR, N1080BBR, N1080EBR, N1080TBR, N6120BV, N620ABR, N620BBR, N620TBR, N640ABR, N640BBR, N640TBR, N660ABR, N660BBR, N660TBR, N680ABR, N680BBR, N680TBR, N7120BV, N720ABR, N720BBR, N720TBR, N740ABR, N740BBR, N740TBR, N760ABR, N760BBR, N760TBR, N780ABR, N780BBR, N780TBR, N8120BV, N820ABR, N820BBR, N820TBR, N840ABR, N840BBR, N840EBR, N840TBR, N860ABR, N860BBR, N860EBR, N860TBR, N880ABR, N880BBR, N880EBR, N880TBR, N920ABR, N920BBR, N920TBR, N940ABR, N940BBR, N940TBR, N960ABR, N960BBR, N960TBR, N980ABR, N980BBR, N980TBR.