FDA Device recall Z-0726-2022
HeartWare Ventricular Assist Device (HVAD) System
- FDA Device
- Class II
- ongoing
- Published 2022-03-16
On 2022-03-16 the FDA classified a Class II recall of HeartWare Ventricular Assist Device (HVAD) System by Heartware, citing updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0726-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-01-27 |
| Published | 2022-03-16 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Heartware |
| Distribution | US |
Product
HeartWare Ventricular Assist Device (HVAD) System
Reason
Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.
Identifiers
| code_info | All lots/serials for the following product model numbers: 1440, 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 14…All lots/serials for the following product model numbers: 1440, 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US, 1440, 1650, 1650, 1650CA-CLIN, 1650DE |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.