FDA Device recall Z-0725-2022

Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722…

On 2022-03-16 the FDA classified a Class II recall of Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722… by Philips North America, citing due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0725-2022
ClassificationClass II
Statusongoing
Initiated2021-11-24
Published2022-03-16
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionUS

Product

Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031

Reason

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Identifiers

code_infoS/N: OUS 109

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.