RecallRadar

FDA Device recall Z-0722-2026

Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02

On 2025-12-03 the FDA classified a Class II recall of Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02 by Djo, citing the IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0722-2026
ClassificationClass II
Statusongoing
Initiated2025-09-25
Published2025-12-03
Last updated2026-09-13 11:57 UTC
CompanyDjo
DistributionUS

Product

Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02

Reason

The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.

Identifiers

code_infoREF/UDI-DI/Lot: DP163CT02/00888912164382/190523, 171221. DP163CT03-BLK-L,XL/00888912164481/150221, 180821

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0722-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.