FDA Device recall Z-0720-2024

Curaplex Field Cric Kit with ET Tube, Item Number 36611

On 2024-01-24 the FDA classified a Class II recall of Curaplex Field Cric Kit with ET Tube, Item Number 36611 by Sarnova Hc, citing it was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0720-2024
ClassificationClass II
Statusongoing
Initiated2023-11-06
Published2024-01-24
Last updated2026-09-13 11:57 UTC
CompanySarnova Hc
DistributionUS

Product

Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.

Reason

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Identifiers

code_infoLot Numbers: ASM0037579 ASM0037631 ASM0037905 ASM0038197

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.