FDA Device recall Z-0720-2020
CLASSIC Radiofrequency Cannula - Curved, 10 cm, SMK Compatible Electrode, REF C-1005-S - Product Usage: Disposable Radiofrequency Cannulae…
- FDA Device
- Class II
- terminated
- Published 2020-01-01
On 2020-01-01 the FDA classified a Class II recall of CLASSIC Radiofrequency Cannula - Curved, 10 cm, SMK Compatible Electrode, REF C-1005-S - Product Usage: Disposable Radiofrequency Cannulae… by Abbott Medical, citing A manufacturing error caused a single lot of 10mm active tip length Cannulas to be labeled as 5mm active tip length cannulas.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0720-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-11-11 |
| Published | 2020-01-01 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Abbott Medical |
| Distribution | US |
Product
CLASSIC Radiofrequency Cannula - Curved, 10 cm, SMK Compatible Electrode, REF C-1005-S - Product Usage: Disposable Radiofrequency Cannulae are provided in either straight, curved, or curved blunt geometries with a stylet inserted in the cannula. Cannulae are insulated except for the active tip. The active tip determines the size of the radiofrequency lesion.
Reason
A manufacturing error caused a single lot of 10mm active tip length Cannulas to be labeled as 5mm active tip length cannulas.
Identifiers
| code_info | UDI (01) 05415067028693(17)220131(10) 6865795 lot 6865795 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0720-2020%22