RecallRadar

FDA Device recall Z-0716-2018

Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage

On 2018-03-07 the FDA classified a Class II recall of Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage by Philips Electronics North America, citing there is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0716-2018
ClassificationClass II
Statusterminated
Initiated2018-01-04
Published2018-03-07
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Reason

There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.

Identifiers

code_info
Allura Xper systems - Model #'s 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058…Allura Xper systems - Model #'s 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722063, 722064, 722065, 722066, 722078, 722079, (added as of 6/25/2018) 1120

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0716-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.