FDA Device recall Z-0716-2016
Cardiovascular Allura Centron
- FDA Device
- Class II
- terminated
- Published 2016-02-10
On 2016-02-10 the FDA classified a Class II recall of Cardiovascular Allura Centron by Philips Electronics North America, citing the use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0716-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-03 |
| Published | 2016-02-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Philips Electronics North America |
| Distribution | US |
Product
Cardiovascular Allura Centron; Model Number: 722400 The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable internal diseases or injuries for: " Dedicated vascular and carotid imaging applications, including diagnostic and interventional procedures. " Cardiac imaging applications including diagnostics, interventional rocedures, pacemaker implantations and electrophysiology (EP). " Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
Reason
The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Identifiers
| code_info | All serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0716-2016%22