RecallRadar

FDA Device recall Z-0715-2018

AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications f…

On 2018-03-07 the FDA classified a Class II recall of AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications f… by Mobius Imaging, citing if the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0715-2018
ClassificationClass II
Statusterminated
Initiated2017-09-26
Published2018-03-07
Last updated2026-09-13 11:54 UTC
CompanyMobius Imaging
DistributionUS

Product

AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).

Reason

If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.

Identifiers

code_infoRequested.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0715-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.