FDA Device recall Z-0714-2014

Draeger Infinity Acute Care System Monitoring Solution

On 2014-01-22 the FDA classified a Class II recall of Draeger Infinity Acute Care System Monitoring Solution by Draeger Medical Systems, citing after a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not acti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0714-2014
ClassificationClass II
Statusterminated
Initiated2013-12-10
Published2014-01-22
Last updated2026-09-13 11:53 UTC
CompanyDraeger Medical Systems
DistributionCA, IL, IN, MA, NH, NY, OK, TX

Product

Draeger Infinity Acute Care System Monitoring Solution

Reason

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Identifiers

code_infowith software versions VG2.0.3 and higher.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.