FDA Device recall Z-0713-2018
SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sec…
- FDA Device
- Class II
- terminated
- Published 2018-03-07
On 2018-03-07 the FDA classified a Class II recall of SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sec… by Siemens Medical Solutions Usa, citing the table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0713-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-10-02 |
| Published | 2018-03-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Reason
The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.
Identifiers
| code_info | Serial Numbers: 108054, 108059, 108062, 108069, 108076, 108077 |
|---|