RecallRadar

FDA Device recall Z-0713-2018

SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sec…

On 2018-03-07 the FDA classified a Class II recall of SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sec… by Siemens Medical Solutions Usa, citing the table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0713-2018
ClassificationClass II
Statusterminated
Initiated2017-10-02
Published2018-03-07
Last updated2026-09-13 11:54 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Reason

The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.

Identifiers

code_infoSerial Numbers: 108054, 108059, 108062, 108069, 108076, 108077

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0713-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.