FDA Device recall Z-0710-2015
PKG, 5MM INSERT, ATRAUMATIC GRASPER FORCEPS, DOUBLE ACTION 45CM, P/N 0250080585 Laparoscopic Manual Instruments are intended for cutting, h…
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of PKG, 5MM INSERT, ATRAUMATIC GRASPER FORCEPS, DOUBLE ACTION 45CM, P/N 0250080585 Laparoscopic Manual Instruments are intended for cutting, h… by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0710-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-17 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Endoscopy |
| Distribution | US |
Product
PKG, 5MM INSERT, ATRAUMATIC GRASPER FORCEPS, DOUBLE ACTION 45CM, P/N 0250080585 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Identifiers
| code_info | 343383 941246 941725 942278 942852 943499 944696 945720 1041625 1042488 1042730 1042738 1042827 1140149 1…343383 941246 941725 942278 942852 943499 944696 945720 1041625 1042488 1042730 1042738 1042827 1140149 1141793 1142618 1143650 1144500 1145003 1145229 1146369 1146656 1242931 1243555 1244994 1245515 1245818 1340539 1341533 1341554 1342765 1343189 1343300 1343410 1343707 1345828 1346105 1346547 1346654 1440433 1442515 1443224 1443605 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0710-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.