FDA Device recall Z-0708-2016

UNIQ FD

On 2016-02-10 the FDA classified a Class II recall of UNIQ FD by Philips Electronics North America, citing the use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0708-2016
ClassificationClass II
Statusterminated
Initiated2015-11-03
Published2016-02-10
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

UNIQ FD; Model Numbers: 722028 The Philips UNIQ is intended for use on human patients to perform: " Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. " Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). " Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures

Reason

The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.

Identifiers

code_infoAll serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.