FDA Device recall Z-0707-2015
PKG, 5MM INSERT, BIOPSY FORCEPS, PUNCH, 45CM, P/N 0250080580 Laparoscopic Manual Instruments are intended for cutting, holding, grasping an…
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of PKG, 5MM INSERT, BIOPSY FORCEPS, PUNCH, 45CM, P/N 0250080580 Laparoscopic Manual Instruments are intended for cutting, holding, grasping an… by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0707-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-17 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Endoscopy |
| Distribution | US |
Product
PKG, 5MM INSERT, BIOPSY FORCEPS, PUNCH, 45CM, P/N 0250080580 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Identifiers
| code_info | 746972 840070 840360 840507 841809 842086 842131 1042715 1140001 1141564 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0707-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.