RecallRadar

FDA Device recall Z-0705-2021

30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF: CH3286

On 2020-12-30 the FDA classified a Class II recall of 30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF: CH3286 by Icu Medical, citing potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exp….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0705-2021
ClassificationClass II
Statusongoing
Initiated2020-08-31
Published2020-12-30
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF: CH3286

Reason

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Identifiers

code_infoLots: 4727479, 4784305

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0705-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.