FDA Device recall Z-0703-2022
QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture)
- FDA Device
- Class II
- terminated
- Published 2022-03-09
On 2022-03-09 the FDA classified a Class II recall of QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture) by Microbiologics, citing the Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test posi….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0703-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-12-01 |
| Published | 2022-03-09 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Microbiologics |
| Distribution | US |
Product
QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture). Sets include: 1. 0864E3 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356
Reason
The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.
Identifiers
| code_info | Catalog Number: 8164; Lot Number (Expiration Date): 8164-45 (12/16/2015), 8164-46 (12/16/2015), 8164-47 (09/02/2016),…Catalog Number: 8164; Lot Number (Expiration Date): 8164-45 (12/16/2015), 8164-46 (12/16/2015), 8164-47 (09/02/2016), 8164-48 (09/02/2016), and 8164-49 (10/03/2016); UDI: 70845357028623 |
|---|