FDA Device recall Z-0702-2020
NATRELLE Resterilizable Sizers for the styles and sizes: Style 15, Style 20, Inspira X (responsive gel), Inspira F (responsive gel), Inspir…
- FDA Device
- Class III
- terminated
- Published 2019-12-25
On 2019-12-25 the FDA classified a Class III recall of NATRELLE Resterilizable Sizers for the styles and sizes: Style 15, Style 20, Inspira X (responsive gel), Inspira F (responsive gel), Inspir… by Allergan, citing the firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0702-2020 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2019-11-06 |
| Published | 2019-12-25 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Allergan |
| Distribution | AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY |
Product
NATRELLE Resterilizable Sizers for the styles and sizes: Style 15, Style 20, Inspira X (responsive gel), Inspira F (responsive gel), Inspira LP (responsive gel), Inspira M (responsive gel), Inspira L (responsive gel), Inspira ST L (Soft Touch gel), Inspira ST LP (Soft Touch Gel), Inspira ST M (Soft Touch Gel), Inspira ST F (Soft Touch Gel), Inspira ST X (Soft Touch Gel) Style 10/UDI: NATRELLE Sty 10-360 Rstrl Sizer Gel Mod 10888628005587 NATRELLE Sty 10-390 Rstrl Sizer Gel Mod 10888628005594 NATRELLE Sty 10-420 Rstrl Sizer Gel Mod 10888628005600 NATRELLE Sty 10-450 Rstrl Sizer Gel Mod 10888628005617 NATRELLE Sty 10-480 Rstrl Sizer Gel Mod 10888628005624 NATRELLE Sty 10-510 Rstrl Sizer Gel Mod 10888628005631 NATRELLE Sty 10-700 Rstrl Sizer Gel Mod 10888628005679 NATRELLE Sty 10-750 Rstrl Sizer Gel Mod 10888628005686 NATRELLE Sty 10-800 Rstrl Sizer Gel Mod 10888628005693 Style 15/UDI: NATRELLE Sty 15-304 Rstrl Sizer Gel Mid 10888628005778 NATRELLE Sty 15-339 Rstrl Sizer Gel Mid 10888628005785 NATRELLE Sty 15-371 Rstrl Sizer Gel Mid 10888628005792 NATRELLE Sty 15-397 Rstrl Sizer Gel Mid 10888628005808 NATRELLE Sty 15-421 Rstrl Sizer Gel Mid 10888628005815 NATRELLE Sty 15-457 Rstrl Sizer Gel Mid 10888628005822 NATRELLE Sty 15-492 Rstrl Sizer Gel Mid 10888628005839 NATRELLE Sty 15-533 Rstrl Sizer Gel Mid 10888628005846 NATRELLE Sty 15-575 Rstrl Sizer Gel Mid 10888628005853 NATRELLE Sty 15-616 Rstrl Sizer Gel Mid 10888628005860 Style 20/UDI: NATRELLE Sty 20-350 Rstrl Sizer Gel High 10888628006003 NATRELLE Sty 20-700 Rstrl Sizer Gel High 10888628006102 NATRELLE Sty 20-750 Rstrl Sizer Gel High 10888628006119 Inspira X (responsive Gel)/UDI: Natrelle Inspira X-Full 205cc Sizer (US) 10888628009431 Natrelle Inspira X-Full 225cc Sizer (US) 10888628009448 Natrelle Inspira X-Full 255cc Sizer (US) 10888628009455 Natrelle Inspira X-Full 285cc Sizer (US) 10888628009462 Natrelle Inspira X-Full 310cc Sizer (US) 10888628009479 Natrelle Inspira X-Full 340cc Sizer (US) 10888628009486 Natrelle Inspira X-Full 375cc Sizer (US) 10888628009493 Natrelle Inspira X-Full 400cc Sizer (US) 10888628009509 Natrelle Inspira X-Full 420cc Sizer (US) 10888628009516 Natrelle Inspira X-Full 445cc Sizer (US) 10888628009523 Natrelle Inspira X-Full 470cc Sizer (US) 10888628009530 Natrelle Inspira X-Full 495cc Sizer (US) 10888628009547 Natrelle Inspira X-Full 525cc Sizer (US) 10888628009554 Natrelle Inspira X-Full 545cc Sizer (US) 10888628009561 Natrelle Inspira X-Full 560cc Sizer (US) 10888628009578 Natrelle Inspira X-Full 580cc Sizer (US) 10888628009585 Natrelle Inspira X-Full 615cc Sizer (US) 10888628009592 Natrelle Inspira X-Full 650cc Sizer (US) 10888628009608 Natrelle Inspira X-Full 700cc Sizer (US) 10888628009615 Natrelle Inspira X-Full 750cc Sizer (US) 10888628009622 Natrelle Inspira X-Full 800cc Sizer (US) 10888628009639 Inspira F (responsive Gel)/UDI: MSZ-F180 Natrelle Inspira FullProf 180cc Sizer US 10888628009226 MSZ-F200 Natrelle Inspira FullProf 200cc Sizer US 10888628009233 MSZ-F220 Natrelle Inspira FullProf 220cc Sizer US 10888628009240 MSZ-F240 Natrelle Inspira FullProf 240cc Sizer US 10888628009257 MSZ-F265 Natrelle Inspira FullProf 265cc Sizer US 10888628009264 MSZ-F295 Natrelle Inspira FullProf 295cc Sizer US 10888628009271 MSZ-F325 Natrelle Inspira FullProf 325cc Sizer US 10888628009288 MSZ-F335 Natrelle Inspira FullProf 335cc Sizer US 10888628009295 MSZ-F345 Natrelle Inspira FullProf 345cc Sizer US 10888628009301 MSZ-F365 Natrelle Inspira FullProf 365cc Sizer US 10888628009318 MSZ-F385 Natrelle Inspira FullProf 385cc Sizer US 10888628009325 MSZ-F415 Natrelle Inspira FullProf 415cc Sizer US 10888628009332 MSZ-F450 Natrelle Inspira FullProf 450cc Sizer US 10888628009349 MSZ-F485 Natrelle Inspira FullProf 485cc Sizer US 10888628009356 MSZ-F520 Natrelle Inspira FullProf 520cc Sizer US 10888628009363 MSZ-F560 Natrelle Inspira FullProf 560cc Sizer US 10888628009370 MSZ-F605 Natrelle
Reason
The firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
Identifiers
| code_info | MFG BATCHES: 2805096 2836786 2884422 2889999 2871740 2878486 2797264 2884426 3036126 3059070 3076451 312445…MFG BATCHES: 2805096 2836786 2884422 2889999 2871740 2878486 2797264 2884426 3036126 3059070 3076451 3124452 3131479 3139882 3154328 3157493 3173220 3176664 3213097 3228743 2776706 2884427 2909055 2921210 3008017 3022733 3035530 3059463 3081093 3096446 3124773 3127774 3142976 3154566 3157086 3157779 3160817 3164079 3167339 3173664 3176918 3182507 3203867 3208283 3211090 3226125 2830860 2868178 2884428 2917776 2958518 2961802 3016093 3019001 3036261 3059268 3076606 3091713 3094037 3124627 3127656 3142812 3148324 3160679 3167175 3189825 3192733 3195360 3210981 3222913 3225690 3228367 2776708 2836794 2884429 2889922 2946579 2954640 2969639 2984574 2996667 3022702 3033348 3036318 3048122 3054598 3059270 3072727 3086874 3094041 3111501 3122044 3124633 3127660 3148328 3157024 3160683 3182357 3185283 3187738 3192736 3194955 3195363 3201094 3210983 3213144 3218861 3220208 3222917 28573 |
|---|