FDA Device recall Z-0702-2017
MANOS EX (CAT#TY-12-101) Product Usage: Carpal Tunnel Release: as a tool for tissue release of the transverse carpal ligament in patients w…
- FDA Device
- Class II
- terminated
- Published 2016-12-14
On 2016-12-14 the FDA classified a Class II recall of MANOS EX (CAT#TY-12-101) Product Usage: Carpal Tunnel Release: as a tool for tissue release of the transverse carpal ligament in patients w… by Thayer Intellectual Property, citing the firm failed to notify current customers of IFU update/warning. During a recent FDA inspection, it was discovered that Thayer Intellectual Property, Inc. had revised their IFU….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0702-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-11-04 |
| Published | 2016-12-14 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Thayer Intellectual Property |
| Distribution | US |
Product
MANOS EX (CAT#TY-12-101) Product Usage: Carpal Tunnel Release: as a tool for tissue release of the transverse carpal ligament in patients with Carpal Tunnel Syndrome who fail conservative therapy. Plantar Fasciotomy: as a tool for tissue recession in patients with plantar fasciitis who fail conservative therapy.
Reason
The firm failed to notify current customers of IFU update/warning. During a recent FDA inspection, it was discovered that Thayer Intellectual Property, Inc. had revised their IFU to include a warning not to bend, deform or otherwise modify any part of the device. The warning was added as a result of a MAUDE report filed in December 2014 where a piece of the MANOS EX broke during use and was left in the patient's hand. The patient suffered a tendon injury and an additional surgery was required to remove the broken piece of the device and repair the tendon. FDA inspection found that while the firm had updated their IFU, they had not notified customers who had been shipped the device with the older version that did not include the warning. Risk: It is important for MANOS EX users to be aware that bending, deforming, or modifying the device could result in device breakage which could impact patient safety and device performance.
Identifiers
| code_info | LOT 1055337, Expiry 4/01/2015; LOT 1071800. Expiry 5/1/2017; LOT 1102580; Expiry 8/1/2017; LOT 1154179; Expiry 12/01/2017. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0702-2017%22