RecallRadar

FDA Device recall Z-0701-2019

IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequen…

On 2019-01-16 the FDA classified a Class III recall of IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequen… by Boston Scientific, citing supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not refle….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0701-2019
ClassificationClass III
Statusongoing
Initiated2018-11-28
Published2019-01-16
Last updated2026-09-13 11:55 UTC
CompanyBoston Scientific
Distributionn/a

Product

IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Reason

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Identifiers

code_info
IntellaNav Open-Irrigated Ablation Catheter Material Number: M004R96200 Batch: 22208397 Expiration Date: 6/4/2021 Ba…IntellaNav Open-Irrigated Ablation Catheter Material Number: M004R96200 Batch: 22208397 Expiration Date: 6/4/2021 Batch: 22389516 Expiration Date: 7/16/2021 Batch: 22525798 Expiration Date: 2/14/2020

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0701-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.