RecallRadar

FDA Device recall Z-0701-2017

3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and 3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case) P…

On 2016-12-14 the FDA classified a Class II recall of 3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and 3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case) P… by 3m Health Care Business, citing 3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product. This may prevent or impede the safe return of electrosurg….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0701-2017
ClassificationClass II
Statusterminated
Initiated2016-10-26
Published2016-12-14
Last updated2026-09-13 11:54 UTC
Company3m Health Care Business
DistributionUS

Product

3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and 3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case) Product Usage: 3M Universal Electrosurgical Pads are designed to work with most electrosurgical unite (ESU) for virtually every surgical application where electrosurgery is utilized to provide a safe return path for electrosurgical current. Solid Universal Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Style Universal Electrosurgical Pad is for use with ESUs that have a CQMS (i.e. REM, ARM, NESSY, etc.). 3M Universal Electrosurgical Pads are designed to be used on any patient where full skin contact and a suitable placement site can be obtained. There are no patient weight restrictions for use of this product. Use of this product for unintended application could lead to an unsafe condition.

Reason

3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product. This may prevent or impede the safe return of electrosurgical current following Electrosurgical Unit (ESU) activation. This defect has the potential to increase the risk of a patient burn directly under the pad or at an alternate site of the body.

Identifiers

code_info
Product REF 9130 Lot Numbers: 2019-02DT (Expiration 02-2019) and 2019-03DU (Expiration 03-2019) Product REF 9735-LP…Product REF 9130 Lot Numbers: 2019-02DT (Expiration 02-2019) and 2019-03DU (Expiration 03-2019) Product REF 9735-LP Lot Numbers: 2019-02DB and 2019-02CY. (Both with Expiration 02-2019)