FDA Device recall Z-0700-2017
Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurosti…
- FDA Device
- Class II
- terminated
- Published 2016-12-14
On 2016-12-14 the FDA classified a Class II recall of Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurosti… by Medtronic Neuromodulation, citing medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 9….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0700-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-10-03 |
| Published | 2016-12-14 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medtronic Neuromodulation |
| Distribution | US |
Product
Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.
Reason
Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.
Identifiers
| code_info | serial numbers beginning with NKA4 or NKU4 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0700-2017%22