FDA Device recall Z-0697-2025

HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4

On 2024-12-25 the FDA classified a Class II recall of HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4 by Medline Industries Northfield, citing complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip w….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0697-2025
ClassificationClass II
Statusongoing
Initiated2024-11-18
Published2024-12-25
Last updated2026-09-13 11:57 UTC
CompanyMedline Industries Northfield
Distributionn/a

Product

HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4; tracheal tube fixation device

Reason

Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.

Identifiers

code_infoUDI/DI 20193489005322 (case), 10193489005325 (each); All Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.