FDA Device recall Z-0697-2013
25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the qu…
- FDA Device
- Class II
- terminated
- Published 2013-01-23
On 2013-01-23 the FDA classified a Class II recall of 25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the qu… by Ids Immunodiagnostic Systems, citing immunodiagnostic Systems is recalling the 25-Hydroxy Vitamin D EIA, AC-57F1 because the QC report was found to have an erroneous calibrator value assigned. The value of 24.4 ng/mL….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0697-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-11-07 |
| Published | 2013-01-23 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ids Immunodiagnostic Systems |
| Distribution | US |
Product
25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the quantitative determination of 25-hydroxyvitamin D (25-OH D) and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.
Reason
Immunodiagnostic Systems is recalling the 25-Hydroxy Vitamin D EIA, AC-57F1 because the QC report was found to have an erroneous calibrator value assigned. The value of 24.4 ng/mL for calibrator 4 was stated on the QC report, however the correct value should have read 28.4 ng/mL.
Identifiers
| code_info | Lot # 17628 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0697-2013%22