FDA Device recall Z-0692-2019
DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in…
- FDA Device
- Class II
- terminated
- Published 2019-01-16
On 2019-01-16 the FDA classified a Class II recall of DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in… by Agfa Gevaert N V, citing after an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0692-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-12-04 |
| Published | 2019-01-16 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Agfa Gevaert N V |
| Distribution | US |
Product
DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.
Reason
After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.
Identifiers
| code_info | Serial Numbers: A5430000015 A5430000020 A5430000030 A5430000042 A5430000052 A5430000061 A5430000066 A5430000067 A5430000079 A5430000081 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0692-2019%22