FDA Device recall Z-0692-2016

Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831

On 2016-02-10 the FDA classified a Class I recall of Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831 by Draeger Medical, citing drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventila….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0692-2016
ClassificationClass I
Statusterminated
Initiated2015-12-22
Published2016-02-10
Last updated2026-09-13 11:54 UTC
CompanyDraeger Medical
DistributionUS

Product

Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831. Volume controlled and pressure-controlled emergency and transport ventilators.

Reason

Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.

Identifiers

code_infoAll Oxylog 3000 Ventilators and Oxylog 3000 Plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0692-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.